RF pulse
射频脉冲
A brief burst of radio waves at the Larmor frequency that tips the protons' net magnetization away from the main field by a chosen flip angle. Every MRI excitation and spin echo starts with one; the heat it deposits is a main MRI safety limit.
The RF coil produces a weak magnetic field , a few microtesla, rotating at the Larmor frequency. Seen from a frame rotating with the spins, the magnetization simply turns about , through a flip angle set by amplitude and duration τ. A 90° pulse lays the magnetization flat in the transverse plane for maximum signal; a 180° pulse inverts it.
In 1950 Erwin Hahn found that two pulses in succession produce an echo: the second pulse makes spins that have fanned out in an uneven field converge again. The spin echo became the basic sequence of clinical MRI for decades, while fast gradient-echo sequences use flip angles well below 90°.
Tissue absorbs RF energy as heat, measured as the specific absorption rate (SAR, W/kg). SAR grows roughly with the square of field strength, so the same sequence deposits about four times the power at 3 T as at 1.5 T, and high-field protocols are often slowed down or use smaller flip angles.
- Hahn EL. Spin echoes. Phys Rev. 1950;80(4):580–594. doi.org/10.1103/PhysRev.80.580
- Spin echo. Wikipedia. en.wikipedia.org/wiki/Spin_echo
Key works · 5
Landmark1950Physical Review7,007 citations
Spin Echoes
Hahn triggered nuclear magnetic resonance in liquids with sequences of radio-frequency pulses. Magnetization that had dephased produced observable echoes at times set by the pulse intervals. Hahn explained this refocusing, measured relaxation from echo amplitudes, and examined how molecular self-diffusion affected those measurements.
SignificanceSpin echoes supplied a way to refocus dephasing caused by static field inhomogeneity and made tissue-dependent transverse relaxation useful for image contrast. Fast spin-echo imaging subsequently used trains of refocused echoes to collect more image data per excitation.
DOI Erwin L. Hahn nuclear magnetic resonanceRF pulseT1 relaxationT2 relaxationspin echo
Landmark1986Journal of Magnetic Resonance1,492 citations
FLASH imaging. Rapid NMR imaging using low flip-angle pulses
FLASH used small flip-angle radio-frequency pulses to retain longitudinal magnetization for the next excitation, allowing successive measurements at short intervals. The sequence reversed the readout gradient to generate an echo. Demonstrated acquisition times were about one second for 64 by 128 pixels and six seconds for 256 by 256.
SignificanceCombining small excitations with gradient echoes established a way to speed MRI by shortening the repetition interval. Kose’s technical history traces the development of fast T1-weighted imaging from FLASH to later sequences that used radio-frequency spoiling to control residual transverse magnetization.
DOI Full text Axel HaaseJens FrahmKlaus-Dietmar Merboldt magnetic resonance imagingRF pulsetemporal resolution
Key1974Journal of Physics C: Solid State Physics370 citations
Image formation in NMR by a selective irradiative process
The authors combined switched magnetic field gradients with frequency-selective radio-frequency irradiation to excite or saturate spins in chosen parts of a sample. They reported preliminary experiments illustrating the components of the method and produced a two-dimensional image of a simple object.
SignificanceApplying a selective pulse in a gradient made it possible to prescribe the slice to be imaged. Spin-warp imaging combined this form of slice selection with frequency and phase encoding, a combination that became standard in clinical two-dimensional MRI.
DOI Allen N. GarrowayPeter Mansfield RF pulsegradient coilmagnetic resonance imaging
Key1992Magnetic Resonance in Medicine1,842 citations
Perfusion imaging
The authors used radio-frequency pulses at 4.7 tesla to saturate water protons in arterial blood flowing toward the rat brain. They compared the resulting images with a distal saturation control to estimate regional cerebral blood flow, and detected changes associated with raised arterial carbon dioxide.
SignificanceThe rat experiment demonstrated quantitative perfusion imaging using arterial water as an endogenous tracer. This measurement principle is central to arterial spin labeling. A 2015 consensus from the ISMRM perfusion group and the European consortium for ASL in dementia gave implementation guidance for clinical MRI.
DOI John A. DetreAlan P. Koretsky magnetic resonance imagingT1 relaxationRF pulse
Key2001Magnetic Resonance in Medicine980 citations
7T vs. 4T: RF power, homogeneity, and signal-to-noise comparison in head images
The authors compared human head imaging at 7 and 4 tesla using measurements and electromagnetic modeling. At 7 tesla, average brain signal-to-noise was at least 1.6 times the value at 4 tesla, while the radio-frequency field was less uniform and excitation at the center of the head required more power.
SignificanceThe controlled comparison quantified both the signal gain and the radio-frequency engineering demands of ultrahigh field, providing a reference for subsequent coil and imaging work. The US Food and Drug Administration cleared the first clinical 7 tesla device in 2017.
DOI J. Thomas VaughanKâmil Uğurbil ultra-high-field MRIteslasignal-to-noise ratioRF pulse