After excitation, the transverse magnetization fades as protons fall out of step with one another, with time constant T2, typically tens of milliseconds and always shorter than T1. Fluid keeps its T2 longest, so edema and effusions appear bright on T2-weighted images, where many lesions first show.
Transverse magnetization decays as , leaving 37% after one T2. Each proton feels a local field slightly disturbed by its neighbors, so spins precess at slightly different rates, their phases fan out, and their signals cancel. At 1.5 T white matter has a T2 of about 70 ms, gray matter about 95 ms, cerebrospinal fluid over a second; unlike T1, T2 hardly changes with field strength.
The observed decay is faster still, because the main field itself is never perfectly uniform; the combined time constant is called T2*. Erwin Hahn's spin echo of 1950 reverses the dephasing caused by static field inhomogeneity but cannot undo true T2 loss, which is how T2 is measured.
With a long repetition time and a long echo time (TE around 80–100 ms), short-T2 tissue has already faded while fluid stays bright. Tumours, inflammation and infarcts usually carry extra water and stand out on T2-weighted images, while T2* sensitivity to iron and blood is the basis of BOLD functional MRI.
- Stanisz GJ, Odrobina EE, Pun J, et al. T1, T2 relaxation and magnetization transfer in tissue at 3T. Magn Reson Med. 2005;54(3):507–512. doi.org/10.1002/mrm.20605
- Hahn EL. Spin echoes. Phys Rev. 1950;80(4):580–594. doi.org/10.1103/PhysRev.80.580
Key works · 5
Landmark1950Physical Review7,007 citations
Spin Echoes
Hahn triggered nuclear magnetic resonance in liquids with sequences of radio-frequency pulses. Magnetization that had dephased produced observable echoes at times set by the pulse intervals. Hahn explained this refocusing, measured relaxation from echo amplitudes, and examined how molecular self-diffusion affected those measurements.
SignificanceSpin echoes supplied a way to refocus dephasing caused by static field inhomogeneity and made tissue-dependent transverse relaxation useful for image contrast. Fast spin-echo imaging subsequently used trains of refocused echoes to collect more image data per excitation.
DOI Erwin L. Hahn nuclear magnetic resonanceRF pulseT1 relaxationT2 relaxationspin echo
Landmark1986Magnetic Resonance in Medicine2,345 citations
RARE imaging: A fast imaging method for clinical MR
The team assigned a different phase-encoding step to each of several spin echoes following excitation, collecting enough data for a 256 by 256 image in 2 to 40 seconds. They examined artifacts caused by changing T2 weighting across the echo train and how phase-encoding order could reduce them.
SignificanceAcquiring multiple k-space lines from a train of refocused echoes became the basis of clinical fast or turbo spin-echo imaging. Later implementations retained this approach to accelerating T2-weighted acquisitions while addressing the blurring and radio-frequency demands of long echo trains.
DOI Jürgen HennigHartmut Friedburg magnetic resonance imagingT2 relaxationk-spacespin echophase encoding
Key1971Science1,911 citations
Tumor Detection by Nuclear Magnetic Resonance
Damadian used spin-echo measurements to compare longitudinal and transverse relaxation in excised samples of six normal rat tissues and two malignant tumors. The two tumor types had relaxation values outside the range of the normal tissues tested, suggesting that the measurements might help distinguish tissue specimens.
SignificanceDifferences among the excised rat tissues gave an early reason to investigate relaxation as a source of contrast. In his Nobel lecture, Lauterbur recounted how Damadian’s result prompted follow-up experiments and how learning about that research preceded his idea of locating signals with a magnetic field gradient.
DOI Raymond Damadian nuclear magnetic resonanceT1 relaxationT2 relaxationspin echo
Key1990Magnetic Resonance in Medicine1,808 citations
Early detection of regional cerebral ischemia in cats: Comparison of diffusion- and T2-weighted MRI and spectroscopy
The authors compared diffusion-weighted imaging, T2-weighted imaging, and magnetic resonance spectroscopy after arterial occlusion in eight cats. Diffusion-weighted images showed increased signal in the ischemic region as early as 45 minutes. T2-weighted spin-echo images did not clearly show injury for two to three hours.
SignificanceThe animal experiment established that diffusion weighting could show changes in acute ischemia earlier than conventional T2 weighting. Subsequent human work tested this early diffusion signal in acute cerebral infarction and helped establish diffusion imaging in stroke assessment.
DOI Michael E. MoseleyPhilip R. Weinstein diffusion MRIT2 relaxation
Key1992Neurology962 citations
Fast magnetic resonance diffusion-weighted imaging of acute human stroke
The authors used rapid diffusion-weighted MRI in patients with cerebral infarction and compared the findings with conventional T2-weighted spin-echo images. Diffusion weighting revealed some infarcts earlier than the conventional sequence, and acute lesions showed lower apparent diffusion coefficients than unaffected regions.
SignificanceThe patient study connected early changes in water diffusion during cerebral ischemia with lesion detection in human stroke. In 1995, the group extended the clinical work to whole-brain echo-planar imaging in 40 patients and confirmed that early diffusion abnormalities could precede changes visible on conventional MRI.
DOI Steven WarachRobert R. Edelman magnetic resonance imagingdiffusion MRIT2 relaxation