T1 relaxation
T1弛豫
After an RF pulse, the protons' magnetization regrows along the main field with time constant T1. Fat recovers fast and water slowly, so on T1-weighted images fat is bright and fluid dark; gadolinium-based contrast agent shortens T1 and makes enhancing lesions light up.
Longitudinal magnetization recovers as , reaching 63% after one T1. Recovery requires the protons to hand energy to surrounding molecules, the "lattice", and works best when molecular motion is near the Larmor frequency. Fat recovers in a few hundred milliseconds; at 1.5 T white matter takes about 0.9 s, gray matter about 1.1 s, and cerebrospinal fluid several seconds.
T1 and T2 entered physics with Felix Bloch's 1946 equations for nuclear magnetization. In 1971 Raymond Damadian found that tumors in rats had markedly longer T1 than normal tissue, the first hint that NMR could diagnose disease.
Scan with a short repetition time (TR, around 500 ms) and slowly recovering tissues never return to full magnetization, so their signal stays low: the image is T1-weighted, good for anatomy. Gadolinium agents shorten the T1 of nearby water, so enhancing tumors and inflammation turn bright. T1 lengthens with field strength; white matter's is about 20% longer at 3 T than at 1.5 T.
- Stanisz GJ, Odrobina EE, Pun J, et al. T1, T2 relaxation and magnetization transfer in tissue at 3T. Magn Reson Med. 2005;54(3):507–512. doi.org/10.1002/mrm.20605
- Damadian R. Tumor detection by nuclear magnetic resonance. Science. 1971;171(3976):1151–1153. doi.org/10.1126/science.171.3976.1151
Key works · 4
Landmark1950Physical Review7,007 citations
Spin Echoes
Hahn triggered nuclear magnetic resonance in liquids with sequences of radio-frequency pulses. Magnetization that had dephased produced observable echoes at times set by the pulse intervals. Hahn explained this refocusing, measured relaxation from echo amplitudes, and examined how molecular self-diffusion affected those measurements.
SignificanceSpin echoes supplied a way to refocus dephasing caused by static field inhomogeneity and made tissue-dependent transverse relaxation useful for image contrast. Fast spin-echo imaging subsequently used trains of refocused echoes to collect more image data per excitation.
DOI Erwin L. Hahn nuclear magnetic resonanceRF pulseT1 relaxationT2 relaxationspin echo
Key1971Science1,911 citations
Tumor Detection by Nuclear Magnetic Resonance
Damadian used spin-echo measurements to compare longitudinal and transverse relaxation in excised samples of six normal rat tissues and two malignant tumors. The two tumor types had relaxation values outside the range of the normal tissues tested, suggesting that the measurements might help distinguish tissue specimens.
SignificanceDifferences among the excised rat tissues gave an early reason to investigate relaxation as a source of contrast. In his Nobel lecture, Lauterbur recounted how Damadian’s result prompted follow-up experiments and how learning about that research preceded his idea of locating signals with a magnetic field gradient.
DOI Raymond Damadian nuclear magnetic resonanceT1 relaxationT2 relaxationspin echo
Key1984The Lancet259 citations
Intravenous chelated gadolinium as a contrast agent in NMR imaging of cerebral tumours
The authors imaged 12 patients with brain tumors before and after intravenous administration of the gadolinium chelate Gd-DTPA. All tumors enhanced after injection. Ring enhancement was the most frequent pattern, and enhancement was visible with both inversion-recovery and spin-echo sequences.
SignificanceThis patient study provided early evidence that a gadolinium chelate could improve visualization of brain tumors. Contrast based on relaxation changes after administration of paramagnetic agents subsequently became an established option for delineating lesions of the central nervous system.
DOI D. H. CarrIan R. Young gadolinium-based contrast agentcontrast agentT1 relaxation
Key1992Magnetic Resonance in Medicine1,842 citations
Perfusion imaging
The authors used radio-frequency pulses at 4.7 tesla to saturate water protons in arterial blood flowing toward the rat brain. They compared the resulting images with a distal saturation control to estimate regional cerebral blood flow, and detected changes associated with raised arterial carbon dioxide.
SignificanceThe rat experiment demonstrated quantitative perfusion imaging using arterial water as an endogenous tracer. This measurement principle is central to arterial spin labeling. A 2015 consensus from the ISMRM perfusion group and the European consortium for ASL in dementia gave implementation guidance for clinical MRI.
DOI John A. DetreAlan P. Koretsky magnetic resonance imagingT1 relaxationRF pulse