beamforming
波束合成
How an ultrasound imaging array focuses electronically: delay the echo received by each element so that signals from one chosen point line up, then add them. That point is reinforced, and echoes from elsewhere partly cancel.
An array transducer carries anywhere from dozens to a few hundred elements. An echo from a point deep in the body reaches each one by a slightly different path; at a soft-tissue sound speed m/s, a 1 mm difference is about 0.65 μs. Delay each channel by its geometric lag and sum, the method called delay-and-sum, and signals from that point add in phase while those from elsewhere fall out of step. On receive the delays are updated continuously with depth (dynamic focusing); on transmit, delays steer and focus the outgoing beam.
In 1968 Jan Somer in the Netherlands described an electronic phased-array sector scanner that swept its beam by changing delays rather than rocking the probe. Linear, curved and cardiac phased arrays all rest on that principle. Modern plane-wave compounding, with beamforming done in software, reaches thousands of frames per second, enough for shear-wave elastography and ultrafast Doppler.
Beamforming assumes sound travels at 1540 m/s everywhere. In fat it is closer to 1450 m/s, so thick fatty layers throw the delays off and smear the focus, one reason ultrasound images of obese patients are harder to read.
- Somer JC. Electronic sector scanning for ultrasonic diagnosis. Ultrasonics. 1968;6(3):153–159. doi.org/10.1016/0041-624X(68)90277-1
- Montaldo G, Tanter M, Bercoff J, Benech N, Fink M. Coherent plane-wave compounding for very high frame rate ultrasonography and transient elastography. IEEE Trans Ultrason Ferroelectr Freq Control. 2009;56(3):489–506. doi.org/10.1109/TUFFC.2009.1067
- Woo J. Phased arrays and Jan C Somer. ob-ultrasound.net. www.ob-ultrasound.net/somers.html
Key works · 4
Key1968Ultrasonics196 citations
Electronic sector scanning for ultrasonic diagnosis
Somer developed an electronically controlled ultrasound sector scanner to overcome the slow refresh of mechanical B-scanning. An array of 21 elements steered pulses across a 90-degree sector and could repeat the scan across that sector about 30 times per second.
SignificanceThe work showed that an array could sweep a beam across a sector repeatedly without moving the probe. A follow-up report applied this electronic scanner to neurological cases, and phased arrays became a means of acquiring real-time cardiac sections.
DOI Jan C. Somer ultrasound imagingtransducerbeamformingB-modetemporal resolution
Key1990Journal of Digital Imaging271 citations
Real time volumetric ultrasound imaging system
The authors used a 289-element two-dimensional phased array to scan a pyramidal volume with pulse echoes. Parallel processing on reception produced approximately eight volume acquisitions per second and also let the system display sections selected from the acquired three-dimensional data.
SignificanceThe prototype combined a two-dimensional array, parallel reception, and section display from acquired volumes to show how rapidly changing anatomy could be scanned in three dimensions. Later real-time three-dimensional echocardiography developed this approach using matrix-array transducers.
DOI Full text Olaf T. von RammStephen W. Smith ultrasound imagingbeamformingtransducertemporal resolutionvoxel
Key2009IEEE Transactions on Ultrasonics, Ferroelectrics, and Frequency Control2,057 citations
Coherent plane-wave compounding for very high frame rate ultrasonography and transient elastography
The authors improved single plane-wave images by coherently summing echoes from transmissions at different angles. Their model predicted image quality comparable to conventional B-mode imaging with roughly one-tenth as many transmissions. Tissue-mimicking phantom experiments supported the prediction, and the paper also demonstrated its application to transient elastography.
SignificanceCoherent compounding provided a way to improve image quality with a limited number of additional transmissions while preserving high frame rates for tasks such as shear wave tracking. Urs and colleagues explicitly credited the method and adapted it to image anatomy and blood flow in the human eye.
DOI Gabriel MontaldoMathias Fink ultrasound imagingB-modebeamformingtemporal resolutionelastography
Key2014IEEE Transactions on Ultrasonics, Ferroelectrics, and Frequency Control1,356 citations
Ultrafast imaging in biomedical ultrasound
This review explained how plane or diverging wave transmission illuminates a broad region and allows many image locations to be reconstructed from the received array signals. It discussed frame rates above a thousand per second and applications to elasticity imaging and blood flow measurements.
SignificanceThe review linked plane- and diverging-wave acquisition to shear wave tracking and blood flow measurement through a shared beamforming framework. Urs and colleagues cited it when introducing the acquisition principles in their subsequent ocular study, which used compounded plane waves to depict anatomy and flow.
DOI Mickaël TanterMathias Fink beamformingultrasound imagingelastographytemporal resolution