helical CT
螺旋CT
The table moves steadily while the X-ray tube spins continuously, so the beam traces a helix through the patient and a whole body region is captured as a volume in one breath-hold. It turned computed tomography from slice-by-slice imaging into true 3D imaging.
Early scanners were tethered by cables: the gantry turned once, stopped, unwound, and the table stepped forward for the next slice. Slip rings let the gantry rotate without stopping while the table moves continuously. Pitch is table travel per rotation divided by beam collimation; a higher pitch scans faster but samples more sparsely.
Because no plane in a helical data set is sampled completely at one table position, the scanner interpolates along the table axis to synthesize full projections for each slice before reconstruction. Willi Kalender and colleagues at Siemens presented single-breath-hold spiral volumetric CT at the RSNA meeting in 1989 and published it in Radiology in 1990.
Covering the whole chest in one breath removed the gaps left when patients breathed differently between slices, and slices could be reconstructed at any position, overlapping if desired. That made multiplanar reconstruction, volume rendering and CT angiography practical.
- Kalender WA, Seissler W, Klotz E, Vock P. Spiral volumetric CT with single-breath-hold technique, continuous transport, and continuous scanner rotation. Radiology. 1990;176(1):181–183. doi.org/10.1148/radiology.176.1.2353088
Key works · 5
Landmark1990Radiology1,159 citations
Spiral volumetric CT with single-breath-hold technique, continuous transport, and continuous scanner rotation
The investigators combined continuous gantry rotation with controlled continuous table motion, producing a helical acquisition. They interpolated the measurements to reconstruct sections at positions chosen after scanning. Clinical examples included lung nodules and contrast-enhanced studies acquired with single-breath-hold volume coverage.
SignificanceHelical acquisition made it possible to cover an organ continuously in one breath hold and choose reconstruction levels afterward. Subsequent multidetector systems retained helical sampling while increasing the volume covered per gantry rotation and the speed of each examination.
DOI Willi A. KalenderPeter Vock helical CTinterpolationmultiplanar reconstruction
Key1999The New England Journal of Medicine814 citations
A Comparison of Virtual and Conventional Colonoscopy for the Detection of Colorectal Polyps
In a prospective study of 100 patients at high risk of colorectal neoplasia, the investigators reconstructed endoluminal views from helical CT and compared the findings with conventional colonoscopy in the same patients. The virtual examination detected a greater proportion of larger polyps than of smaller ones.
SignificanceComparing CT endoluminal display with conventional colonoscopy in the same patients moved virtual endoscopy from feasibility testing to clinical assessment by lesion size. Subsequent research extended the evaluation from a high-risk study population to asymptomatic people undergoing screening.
DOI Helen M. FenlonJoseph T. Ferrucci computed tomographyhelical CTendoscopysurface renderingCT colonography
Key2000Radiology561 citations
Four Multidetector-Row Helical CT: Image Quality and Volume Coverage Speed
This study compared four- and single-detector-row helical CT using phantom measurements, models, and ratings of clinical images. The investigators measured section profiles, noise, and artifacts. The four-row system achieved two to three times the volume coverage speed at comparable diagnostic image quality.
SignificanceThe comparison provided an early quantitative baseline for the advantages of acquiring several detector rows together. Subsequent multidetector systems extended the same approach to faster coverage of organs while preserving thin sections for multiplanar reformats.
DOI Hui HuStanley H. Fox multidetector CThelical CTspatial resolution
Key2005Medical Physics300 citations
Image reconstruction and image quality evaluation for a 64-slice CT scanner with z-flying focal spot
A CT scanner alternated its focal spot between two longitudinal positions, interleaving the samples from 32 detector rows to acquire 64 overlapping sections per rotation. The study measured longitudinal resolution, section profiles, and noise and evaluated helical image quality using phantom and patient scans.
SignificanceThe scanner demonstrated that geometry could increase longitudinal sampling density beyond the physical detector row count. The technique was implemented in a commercial 64-section system, supporting thin-section volume acquisitions and reformats through the resulting data in multiple planes.
DOI Thomas G. FlohrCynthia H. McCollough multidetector CThelical CTspatial resolution
Key2007Medical Physics1,028 citations
A three-dimensional statistical approach to improved image quality for multislice helical CT
The method used a three-dimensional forward model of multislice helical acquisition, statistical weighting of measurements, and a prior on the image. Phantom tests demonstrated improved resolution and lower noise alongside reduced helical cone-beam artifacts. Reconstructions of patient scans illustrated clinical image appearance.
SignificanceThe work demonstrated model-based iterative reconstruction on real multislice helical CT data. Later evaluations of commercial model-based reconstruction cite this study when examining how spatial resolution and image noise vary with exposure and reconstruction settings.
DOI Full text Jean-Baptiste ThibaultJiang Hsieh iterative reconstructionhelical CTmultidetector CTnoise