dynamic range

动态范围

The ratio between the strongest and weakest signals a system can record or show at once. A detector's dynamic range decides how forgiving it is of exposure errors; a display's narrow range is why images are viewed through a window width & level setting.

Dynamic range can be stated as a ratio or in bits; 1000:1 is about 10 bits. X-ray film has a characteristic curve that is straight only in the middle: a little too much exposure and the film is black, a little too little and it is clear, so radiographers had to set exposures precisely or repeat the shot.

The imaging plates of computed radiography, introduced in the 1980s, and later flat-panel detectors respond roughly linearly over about four orders of magnitude. Exposure errors can be corrected in software at display time, and fewer images need repeating. The flip side is dose creep: an overexposed image still looks fine, so the excess goes unnoticed.

Ultrasound echoes vary in strength by several orders of magnitude and are log-compressed to fit the screen; the 4,096 levels of a CT image are compressed for display with a window setting or a histogram transform.

Sources
  1. Rowlands JA. The physics of computed radiography. Phys Med Biol. 2002;47(23):R123–R166. doi.org/10.1088/0031-9155/47/23/201