window width & level
窗宽窗位
The rule that maps CT numbers to screen gray: the level sets the center, the width sets the range, and everything outside is pure black or pure white. The same data show lung texture in one window and gray–white matter in another.
With level and width , CT numbers below show as black, those above as white, and the range between is spread linearly across the grayscale. CT data are usually stored with 12 bits, over four thousand levels, while a display offers about 256. A brain window of , stretches 0–80 HU across the whole gray scale, about 0.3 HU per gray step, enough to separate gray and white matter that differ by ten or so units. A lung window near , brings out vessels and small nodules in air-filled lung.
The DICOM standard carries a suggested setting with every image, in the Window Center (0028,1050) and Window Width (0028,1051) attributes.
With the wrong window, a finding is displayed as uniform black or white: a rib fracture is invisible in a brain window, ground-glass opacity in a soft-tissue window. Radiologists flick between presets as they read, and the same choice matters when CT is fed to a neural network: the window used in preprocessing determines which range of CT numbers remains distinguishable in the model's input.
- Goldman LW. Principles of CT and CT technology. J Nucl Med Technol. 2007;35(3):115–128. tech.snmjournals.org/content/35/3/115
- DICOM PS3.3, C.11.2 VOI LUT Module (Window Center 0028,1050; Window Width 0028,1051). dicom.nema.org/medical/dicom/current/output/chtml/part03/sect_C.