Image basics

grayscale

灰度

Showing one number at each point as a brightness between black and white. Most medical images measure a single physical quantity per point, so they are naturally grayscale; which tissue looks white depends on what was measured.

Bone is white on a radiograph because of film: bone stops X-rays, the film under it is barely exposed, develops clear, and glows on the lightbox. Digital imaging kept the convention, and DICOM even names both directions: in MONOCHROME2 higher values are brighter, in MONOCHROME1 darker. In MRI, white only means strong signal; fat is bright on T1-weighted images, water on T2.

The eye does not perceive equal steps of luminance as equal. The Grayscale Standard Display Function in Part 14 of DICOM, built on Peter Barten's model of human contrast sensitivity, divides 0.05–4000 cd/m² into 1,023 just-noticeable differences, so that different monitors render the same value as the same perceived gray.

Colour is added only when a second kind of information must sit on top, such as uptake in PET or flow direction in Doppler ultrasound.

Sources
  1. NEMA. DICOM PS3.14: Grayscale Standard Display Function. dicom.nema.org/medical/dicom/current/output/html/part14.html

Key works · 3

Landmark19581957 Eastern Joint Computer Conference126 citations

Experiments in processing pictorial information with a digital computer

R. A. Kirsch, L. Cahn … G. H. Urban

The National Bureau of Standards team used a scanner to convert pictures to discrete samples and passed those values to the SEAC computer. They reported examples with photographs and printed patterns, showing how programmed operations could manipulate and analyze structure within digital images.

SignificanceThe work demonstrated a complete process from a physical picture to an array that a computer could manipulate. NIST identifies the 1957 scanning demonstration as a starting point for digital image processing, and medical image formats subsequently used discrete arrays for image storage.

Key1991Communications of the ACM4,452 citations

The JPEG still picture compression standard

Gregory K. Wallace

Wallace explained JPEG compression for continuous-tone still pictures, including lossy coding based on the discrete cosine transform and quantization and lossless coding based on prediction. He outlined the baseline coding steps in the 1991 ACM article and revisited them in the revised IEEE version published in 1992.

SignificanceThe article became an early reference for JPEG's coding modes and practical use. DICOM subsequently defined transfer syntaxes for encapsulating JPEG-compressed pixel data within medical image objects, allowing systems that support the same syntax to exchange those data.

Notable2007Journal of Digital Imaging209 citations

Increasing the Number of Gray Shades in Medical Display Systems—How Much is Enough?

Tom Kimpe, Tom Tuytschaever

The authors examined grayscale display using human luminance discrimination, the DICOM grayscale display function, and the characteristics of medical LCDs. Under the modeled conditions, increasing the panel's inherent bit depth from 11 to 12 bits yielded only a small improvement in calibration accuracy.

SignificanceThe study distinguished stored bit depth, display luminance levels, and perceptible gray shades. Bender and colleagues cited it in a study that measured just-noticeable luminance steps for 8-, 10-, and 11-bit lookup tables and asked radiologists to compare images displayed with the 8- and 11-bit tables.