Image basics

contrast

对比度

The difference in signal between a structure and its surroundings. Whether a lesion can be seen depends first on how much it differs from normal tissue in the measured quantity; without contrast, resolution does not help.

A common definition is the relative difference . Contrast is built in layers: the natural difference between tissues in some physical quantity, the detector's ability to record it, and the window width & level setting that stretches it on screen.

Each modality measures a different physical property, so tissue contrast differs from one modality to another. X-ray contrast comes from attenuation, so bone stands out from soft tissue while liver and a liver tumor look almost alike; hence iodinated contrast agents, whose K-edge near 33 keV makes them more conspicuous at lower tube voltage. MRI contrast is adjustable: the same tissue, scanned with different TR and TE, can be made T1-, T2- or proton density-weighted.

Contrast only means something next to noise. A lesion a few per cent brighter than its background vanishes when the image is grainy, which is why visibility is judged by the contrast-to-noise ratio.

Key works · 3

Landmark1915United States Patent Office

Method of and apparatus for projecting Röntgen images

Gustav Bucky

Bucky proposed putting an absorbing grid between the object and the photographic plate to intercept secondary radiation. The openings were aligned with rays from the source, allowing the primary beam to reach the plate while preferentially blocking rays traveling at other angles.

SignificanceThe grid established a way to improve radiographic contrast by rejecting scattered radiation before it reached the receptor. Potter's subsequent use of moving parallel strips reduced the visibility of the grid on the radiograph and led to the Potter–Bucky grid used in clinical systems.

Key1921Archives of Radiology and Electrotherapy5 citations

The Bucky Diaphragm Principle Applied to Roentgenography

Hollis E. Potter

Potter replaced the crosswise structure of Bucky's grid with parallel metal strips and moved the grid across the plate during exposure. Uniform motion spread the strip shadows across the plate while the strips still intercepted scattered radiation. Spine radiographs illustrate the resulting suppression of grid lines.

SignificanceThe moving strips addressed a practical obstacle to routine scatter rejection in projection radiography. The Science Museum Group documents the commercial introduction of the Potter–Bucky device in 1921 and holds an example of the equipment used in radiology.

Key1948Journal of the Optical Society of America1,056 citations

The Sensitivity Performance of the Human Eye on an Absolute Scale

Albert Rose

Rose used a detector limited by random photon fluctuations as a benchmark for human visual sensitivity. A television pickup experiment related target visibility to luminance, contrast, and angular size. Under the experimental model he estimated a threshold signal-to-noise ratio of about five.

SignificanceThe result became the Rose criterion used in image quality analysis. Medical physics texts use it as an approximate detectability benchmark for targets in noisy images, where visibility also depends on target size, background, and the viewing task.