Physics

anti-scatter grid

滤线栅

An anti-scatter grid sits between the patient and the detector. It transmits much of the primary X-ray beam and absorbs some obliquely traveling scatter, improving image contrast.

In radiography, photons deflected by Compton scattering can reach the detector along a different path. They add signal that reduces differences between tissues. An anti-scatter grid has thin lead strips separated by channels aligned with the expected paths from the source. Photons traveling at a substantial angle to the channels are more likely to strike the lead.

The grid also absorbs some primary radiation. Maintaining the same detector exposure can require more X-ray output, which ties the improvement in contrast to a dose trade-off. A tilted or incorrectly centered grid can block additional primary photons. This reduces the recorded signal in the affected regions of the image.

Sources
  1. Dance DR, Christofides S, Maidment ADA, McLean ID, Ng KH, eds. Diagnostic Radiology Physics: A Handbook for Teachers and Students. IAEA, 2014. Chapter 6. www-pub.iaea.org/MTCD/Publications/PDF/Pub1564webNew-74666420.pd

Key works · 2

Landmark1915United States Patent Office

Method of and apparatus for projecting Röntgen images

Gustav Bucky

Bucky proposed putting an absorbing grid between the object and the photographic plate to intercept secondary radiation. The openings were aligned with rays from the source, allowing the primary beam to reach the plate while preferentially blocking rays traveling at other angles.

SignificanceThe grid established a way to improve radiographic contrast by rejecting scattered radiation before it reached the receptor. Potter's subsequent use of moving parallel strips reduced the visibility of the grid on the radiograph and led to the Potter–Bucky grid used in clinical systems.

Key1921Archives of Radiology and Electrotherapy5 citations

The Bucky Diaphragm Principle Applied to Roentgenography

Hollis E. Potter

Potter replaced the crosswise structure of Bucky's grid with parallel metal strips and moved the grid across the plate during exposure. Uniform motion spread the strip shadows across the plate while the strips still intercepted scattered radiation. Spine radiographs illustrate the resulting suppression of grid lines.

SignificanceThe moving strips addressed a practical obstacle to routine scatter rejection in projection radiography. The Science Museum Group documents the commercial introduction of the Potter–Bucky device in 1921 and holds an example of the equipment used in radiology.