contrast-to-noise ratioCNR

对比噪声比

The signal difference between two tissues divided by the noise. It predicts whether a lesion can be seen better than contrast or SNR alone, and is the usual yardstick when comparing protocols or cutting dose.

It is defined as . Take liver at 60 HU and a lesion at 50 HU. With 5 HU of noise the CNR is 2 and the lesion is faintly visible; with 15 HU it falls below 1 and the lesion dissolves into the grain. Size matters too: the larger the lesion, the lower the CNR the eye can work with.

There are two ways to raise CNR: widen the difference or lower the noise. The first uses contrast agents, lower CT tube voltage to make iodine brighter, or different TR and TE in MRI. The second uses more dose or time, or denoising and iterative reconstruction.

When a low-dose CT protocol is evaluated, lesion CNR is a common way to compare it with the standard one, more objective than asking whether the images look good. But CNR captures only the average difference, not noise texture or edge sharpness, so reader studies still have the last word.

Key works · 2

Landmark2022Radiology652 citations

First Clinical Photon-counting Detector CT System: Technical Evaluation

Kishore Rajendran, Martin Petersilka … Thomas G. Flohr

The authors characterized a clinical dual-source photon-counting CT scanner with phantom measurements and example examinations from four participants. They measured spatial resolution, noise, section response, and iodine quantitation. The high-resolution configuration achieved a maximum measured in-plane resolution of 125 micrometers.

SignificanceThe evaluation documented performance at the clinical introduction of a commercial photon-counting CT system. A contemporaneous technical baseline for high-resolution imaging and spectral quantitation allowed later assessments, including the Radiology milestones review, to compare these capabilities with earlier CT technology.

Key1948Journal of the Optical Society of America1,056 citations

The Sensitivity Performance of the Human Eye on an Absolute Scale

Albert Rose

Rose used a detector limited by random photon fluctuations as a benchmark for human visual sensitivity. A television pickup experiment related target visibility to luminance, contrast, and angular size. Under the experimental model he estimated a threshold signal-to-noise ratio of about five.

SignificanceThe result became the Rose criterion used in image quality analysis. Medical physics texts use it as an approximate detectability benchmark for targets in noisy images, where visibility also depends on target size, background, and the viewing task.