positron emission tomographyPET

正电子发射断层成像

Imaging a positron-emitting tracer, most often FDG, by detecting the pair of 511 keV photons that fly apart when each positron annihilates. The result is a 3D map of metabolic activity, central to staging cancer.

Isotopes such as fluorine-18 emit positrons, which travel on average less than a millimeter before meeting an electron and annihilating into two 511 keV photons flying off almost exactly back to back. When a ring of detectors registers two photons within a few nanoseconds, the annihilation must lie on the line joining them, a line of response; hundreds of millions of such lines are reconstructed into an image. Because this coincidence trick replaces the lead collimator, PET is far more sensitive than single-photon emission computed tomography.

In 1975 Michel Ter-Pogossian, Michael Phelps and Edward Hoffman at Washington University in St Louis described PETT, their positron-emission transaxial tomograph. The following year FDG was first given to people, letting PET measure glucose metabolism.

Many tumors take up glucose avidly, so FDG-PET can reveal metastases throughout the body. Inflammation, the brain and heart muscle take up FDG too, though, so the images must be read against anatomy, which is why PET/CT became the standard.

Sources
  1. Ter-Pogossian MM, Phelps ME, Hoffman EJ, Mullani NA. A positron-emission transaxial tomograph for nuclear imaging (PETT). Radiology. 1975;114(1):89–98. doi.org/10.1148/114.1.89

Key works · 9

Landmark1975Radiology1,158 citations

A Positron-Emission Transaxial Tomograph for Nuclear Imaging (PETT)

Michel M. Ter-Pogossian, Michael E. Phelps … Nizar A. Mullani

The PETT combined a hexagonal array of 24 sodium iodide detectors with coincidence electronics and computerized cross-sectional reconstruction. Simulations, phantoms, and animal studies were used to evaluate its ability to recover tracer activity and compare its contrast and resolution with scintillation cameras.

SignificancePETT put multiangle coincidence measurement and computerized emission tomography into a single instrument. Later PET histories identify it as a precursor to clinical scanners that use detector rings and quantitative reconstruction to map tracer distributions.

Landmark1979Circulation Research1,368 citations

The [18F]fluorodeoxyglucose method for the measurement of local cerebral glucose utilization in man

M. Reivich, D. Kuhl … T. Ido

Reivich and colleagues combined arterial activity measurements with brain emission tomography after injection of fluorine-18 FDG. Their kinetic model used FDG as a tracer for glucose exchange between plasma and brain and for phosphorylation in tissue. They used the method to estimate regional glucose utilization in two healthy volunteers.

SignificanceThe study brought fluorine-18 FDG into quantitative human imaging and made regional brain metabolism accessible in vivo. Later work tested and refined the kinetic model, while FDG imaging also became widely used for assessing tumor metabolism.

Landmark1982IEEE Transactions on Medical Imaging5,027 citations

Maximum Likelihood Reconstruction for Emission Tomography

L. A. Shepp, Y. Vardi

Shepp and Vardi modeled unobserved emissions as Poisson variables and specified the probabilities that emissions from each voxel would be registered by each detector. Their expectation-maximization algorithm iteratively estimated the underlying emission distribution by maximizing the likelihood of the measured counts.

SignificanceThe model brought counting statistics and detection probabilities directly into emission reconstruction. Maximum-likelihood expectation-maximization became a basis for iterative PET and SPECT, including subsequent ordered-subsets methods that sought to reduce the cost of reaching a useful image.

Landmark1986IEEE Transactions on Nuclear Science553 citations

A Multicrystal Two Dimensional BGO Detector System for Positron Emission Tomography

M. E. Casey, R. Nutt

The authors coupled an array of bismuth germanate scintillation crystals to four photomultiplier tubes through a light guide. The electronics decoded the position of each event from the shared optical signals. This allowed a PET detector to identify individual crystals in two dimensions with a small number of readout channels.

SignificanceThe block detector let PET scanners use many small crystals without dedicating a photomultiplier to each crystal, making finer spatial sampling practical. The first commercial scanner based on detector blocks was installed at Hammersmith and could read out multiple transaxial planes.

Landmark1994IEEE Transactions on Medical Imaging4,178 citations

Accelerated image reconstruction using ordered subsets of projection data

H. M. Hudson, R. S. Larkin

Hudson and Larkin split projection data into ordered subsets and used each subset to perform an expectation-maximization update before moving to the next. A full iteration visits every subset. This accelerated the early progress of image reconstruction in the experiments they reported.

SignificanceOrdered-subsets expectation-maximization reduced the time required to reach useful image quality, making statistical reconstruction more practical in clinical nuclear medicine. IAEA guidance identifies it as a commonly used reconstruction approach for quantitative PET examinations in clinical practice.

Landmark2005Alzheimer's & Dementia1,399 citations

Ways toward an early diagnosis in Alzheimer's disease: The Alzheimer's Disease Neuroimaging Initiative (ADNI)

Susanne G. Mueller, Michael W. Weiner … William Jagust

The paper described ADNI's plan to follow participants with normal cognition, mild cognitive impairment, or early Alzheimer's disease. Repeated MRI, PET, fluid biomarkers, and clinical assessments were to be combined to investigate markers for identifying and monitoring early Alzheimer’s disease across visits.

SignificanceADNI provided a way to examine longitudinal imaging and clinical measures together across participating sites. Its data portal supports research access, enabling tests of candidate biomarkers across modalities or visits from the same study participants.

Key1999Physics in Medicine & Biology716 citations

Calculation of positron range and its effect on the fundamental limit of positron emission tomography system spatial resolution

C. S. Levin, E. J. Hoffman

The authors simulated positron trajectories for several radionuclides and estimated how displacement before annihilation contributes to image blur. They combined that contribution with detector element size and photon noncollinearity to assess achievable resolution for PET systems with different detector dimensions.

SignificanceThe analysis quantified the contributions of isotope physics and scanner geometry to PET resolution. It gives designers and evaluators a reference for judging how much finer detectors can improve resolution before other physical effects dominate.

Key2004Physics in Medicine & Biology2,510 citations

GATE: a simulation toolkit for PET and SPECT

S. Jan, G. Santin … D. Autret

GATE wrapped Geant4 in a scripted toolkit for simulating PET and SPECT measurements. It let users describe detector or source motion and tracer decay over time. The paper provided benchmark examples for installation and validation and discussed comparison with real scanners.

SignificanceA shared simulation toolkit made it possible for groups to compare detectors, acquisition protocols, and correction algorithms under specified conditions. The OpenGATE collaboration continued to maintain the toolkit and subsequently extended it to additional imaging and dosimetry applications.

Key2009Journal of Nuclear Medicine4,103 citations

From RECIST to PERCIST: Evolving Considerations for PET Response Criteria in Solid Tumors

Richard L. Wahl, Heather Jacene … Martin A. Lodge

The paper reviewed anatomical and metabolic response assessment and proposed PERCIST 1.0 for solid tumors. It specified consistent imaging conditions and a reference tissue for comparing scans before and after systemic therapy. Peak lesion FDG uptake was normalized to lean body mass and measured in a small region of fixed volume.

SignificanceThe proposal supplied reproducible rules for evaluating metabolic response across treatment studies. EANM procedure guidelines cite PERCIST when discussing quantitative FDG assessment, and clinical studies have compared its response classifications with other established treatment response criteria.