annihilation
湮灭
When a positron meets an electron, both vanish and their mass becomes two gamma photons of 511 keV each, flying off almost exactly back to back. A PET scanner detects both at once on its detector ring, which locates the event on the line joining them.
511 keV is the electron's rest-mass energy, . The pair is nearly at rest when it annihilates, so momentum conservation sends the two photons in opposite directions. When a PET ring records two 511 keV photons within a window of a few nanoseconds, it logs a coincidence; the line between the two detectors is a line of response, and millions of them are reconstructed into a map of the tracer.
This electronic collimation needs no lead collimator, which makes PET far more sensitive than a gamma camera. Measure the difference in arrival times as well and the event can be placed along the line: a timing resolution of 200 ps corresponds to about 3 cm, the principle of time-of-flight PET.
The photons are not perfectly collinear; the angle deviates by about ±0.25°, which on an 80 cm ring blurs positions by nearly 2 mm. Together with positron range this sets a physical floor on PET resolution. The photons also undergo Compton scattering and attenuation in the body, each of which must be corrected.
- Moses WW. Fundamental limits of spatial resolution in PET. Nucl Instrum Methods Phys Res A. 2011;648(Suppl 1):S236–S240. doi.org/10.1016/j.nima.2010.11.092
- Positron emission tomography. Wikipedia. en.wikipedia.org/wiki/Positron_emission_tomography
Key works · 3
Landmark1975Radiology1,158 citations
A Positron-Emission Transaxial Tomograph for Nuclear Imaging (PETT)
The PETT combined a hexagonal array of 24 sodium iodide detectors with coincidence electronics and computerized cross-sectional reconstruction. Simulations, phantoms, and animal studies were used to evaluate its ability to recover tracer activity and compare its contrast and resolution with scintillation cameras.
SignificancePETT put multiangle coincidence measurement and computerized emission tomography into a single instrument. Later PET histories identify it as a precursor to clinical scanners that use detector rings and quantitative reconstruction to map tracer distributions.
DOI Michel M. Ter-PogossianMichael E. PhelpsEdward J. HoffmanNizar A. Mullani positron emission tomographyannihilationscintillator
Key1955JAMA94 citations
Localization of Intracranial Lesions by Scanning with Positron-Emitting Arsenic
Sweet and Brownell examined intracranial lesions using a positron-emitting arsenic tracer and opposed detectors operating in coincidence. They used the spatial pattern of counts to localize abnormal uptake and illustrated how external measurements could inform the localization of lesions inside the head.
SignificanceThe clinical series showed that positron coincidence measurements could be made in patients and used for lesion localization. Tomographic PET later used coincidences collected from many more angles to reconstruct the internal activity distribution in cross-section.
DOI William H. SweetGordon L. Brownell radiotracerpositronannihilation
Key1999Physics in Medicine & Biology716 citations
Calculation of positron range and its effect on the fundamental limit of positron emission tomography system spatial resolution
The authors simulated positron trajectories for several radionuclides and estimated how displacement before annihilation contributes to image blur. They combined that contribution with detector element size and photon noncollinearity to assess achievable resolution for PET systems with different detector dimensions.
SignificanceThe analysis quantified the contributions of isotope physics and scanner geometry to PET resolution. It gives designers and evaluators a reference for judging how much finer detectors can improve resolution before other physical effects dominate.
DOI Craig S. LevinEdward J. Hoffman positronannihilationpositron emission tomographyspatial resolution