Michael E. Casey

Co-developed the PET block detector that reads a scintillator array with four photomultiplier tubes.

Nuclear medicine and molecular imaging OpenAlex

Computer Technology and Imaging (1986)

Key works · 1

Landmark1986IEEE Transactions on Nuclear Science553 citations

A Multicrystal Two Dimensional BGO Detector System for Positron Emission Tomography

M. E. Casey, R. Nutt

The authors coupled an array of bismuth germanate scintillation crystals to four photomultiplier tubes through a light guide. The electronics decoded the position of each event from the shared optical signals. This allowed a PET detector to identify individual crystals in two dimensions with a small number of readout channels.

SignificanceThe block detector let PET scanners use many small crystals without dedicating a photomultiplier to each crystal, making finer spatial sampling practical. The first commercial scanner based on detector blocks was installed at Hammersmith and could read out multiple transaxial planes.