Landmark1986IEEE Transactions on Nuclear Science553 citations
A Multicrystal Two Dimensional BGO Detector System for Positron Emission Tomography
The authors coupled an array of bismuth germanate scintillation crystals to four photomultiplier tubes through a light guide. The electronics decoded the position of each event from the shared optical signals. This allowed a PET detector to identify individual crystals in two dimensions with a small number of readout channels.
SignificanceThe block detector let PET scanners use many small crystals without dedicating a photomultiplier to each crystal, making finer spatial sampling practical. The first commercial scanner based on detector blocks was installed at Hammersmith and could read out multiple transaxial planes.
DOI Michael E. CaseyRonald Nutt positron emission tomographyscintillatorspatial resolution