MG·LB · Researchers

Bruce H. Hasegawa

Led development of prototypes combining X-ray CT and single-photon emission tomography.

Nuclear medicine and molecular imagingCT imaging and reconstruction OpenAlex

University of California San Francisco

Key works · 1

Key1996IEEE Transactions on Nuclear Science196 citations

Attenuation correction of SPECT using X-ray CT on an emission-transmission CT system: myocardial perfusion assessment

S. C. Blankespoor, X. Xu … B. H. Hasegawa

The researchers registered data from juxtaposed CT and SPECT scanners and converted the CT measurements to attenuation maps at 140 keV. Attenuation correction reduced underestimation of myocardial activity in a torso phantom and in pigs. An additional partial-volume correction improved agreement with tissue samples in pigs but overestimated activity in the phantom.

SignificanceThis study quantified the benefit and limitations of using registered CT data for attenuation correction on a combined emission–transmission system. UCSF traces its human-scale SPECT/CT prototype for clinical research to the group's work on registered emission and transmission imaging.