Modalities · 2011

PET/MR

positron emission tomography detectors built into an magnetic resonance imaging scanner so both run at once. It pairs metabolic imaging with MRI's superior soft-tissue contrast and avoids CT's radiation, but the machines are expensive and correcting for attenuation is harder.

The obstacle is the magnet. Conventional PET reads its scintillation light with photomultiplier tubes, which fail in strong magnetic fields. Integrated PET/MR uses semiconductor light sensors that are insensitive to the field, such as avalanche photodiodes, and squeezes the detector ring between the magnet and the gradient coils without disturbing the MR system's radio frequencies or field uniformity.

Siemens' Biograph mMR, described in 2011 as the first commercially available integrated whole-body PET/MR scanner, used exactly such avalanche photodiodes.

PET/MR suits cases that need soft-tissue detail with less radiation: children with cancer, neurological disease, prostate and pelvic tumors. Its weak point is attenuation correction. MR signal reflects protons, not electron density, and bone and air are both nearly invisible on routine sequences, so attenuation maps must be estimated by segmentation, atlases or deep learning.

Sources
  1. Delso G, et al. Performance measurements of the Siemens mMR integrated whole-body PET/MR scanner. J Nucl Med. 2011;52(12):1914–1922. doi.org/10.2967/jnumed.111.092726

Key works · 2

Key2011Physics in Medicine & Biology439 citations

Design and performance evaluation of a whole-body Ingenuity TF PET–MRI system

H. Zaidi, N. Ojha … P. Maniawski

This study evaluated a sequential whole-body system combining a time-of-flight PET scanner with 3 T MRI. The PET photomultipliers received magnetic shielding and the electronics were relocated to reduce mutual interference. Measurements characterized PET performance and the use of MR-derived attenuation maps.

SignificanceThe Philips sequential design retained photomultiplier-based PET detectors through shielding and physical separation. In their evaluation of the simultaneous Siemens scanner, Delso and colleagues discussed the Philips design as a contemporaneous clinical alternative. Their standardized phantom measurements characterized the Siemens scanner.

Key2011Journal of Nuclear Medicine960 citations

Performance Measurements of the Siemens mMR Integrated Whole-Body PET/MR Scanner

Gaspar Delso, Sebastian Fürst … Sibylle I. Ziegler

Delso and colleagues measured an integrated whole-body PET/MR scanner while each subsystem operated independently and while both acquired data together. PET measurements included resolution, sensitivity, scatter, and count losses. MRI quality-control tests covered geometric accuracy and signal quality. The authors reported no perceptible effect on overall performance from integration under the conditions they tested.

SignificanceThe evaluation documented the first commercial integrated whole-body PET/MR scanner with simultaneous acquisition. Collecting emissions and MR data over the same period made it possible to investigate tracer behavior alongside MR measures of anatomy and function.