MG·LB · Researchers

Ramsey D. Badawi

Co-developed the EXPLORER total-body PET scanner and reported its first human images.

Nuclear medicine and molecular imaging ORCID OpenAlex

University of California Davis (2019)

Key works · 2

Landmark2019Journal of Nuclear Medicine778 citations

First Human Imaging Studies with the EXPLORER Total-Body PET Scanner

Ramsey D. Badawi, Hongcheng Shi … Patricia M. Price

This report presented initial human imaging with the EXPLORER PET/CT scanner. Its 194 cm axial field of view allowed activity to be recorded across the body at one bed position. Examples illustrated several ways of using the added sensitivity, including rapid, low-activity, delayed, and dynamic whole-body studies.

SignificanceSimultaneous coverage made it possible to observe tracer kinetics in multiple organs in the same time frame. Subsequent characterization of the instrument measured sensitivity, count-rate performance, and image quality, creating benchmarks for later long-axial-field systems.

Key2018Journal of Nuclear Medicine827 citations

Total-Body PET: Maximizing Sensitivity to Create New Opportunities for Clinical Research and Patient Care

Simon R. Cherry, Terry Jones … Ramsey D. Badawi

The authors reviewed geometric and timing improvements that could increase PET sensitivity. They estimated that extending axial coverage to the whole body could increase effective sensitivity about 40-fold for whole-body imaging and discussed the implications for scan duration, administered activity, and dynamic studies.

SignificanceThe article provided a quantitative rationale for a scanner with an extended axial field and distinguished the gains for whole-body imaging from those for a single organ. The EXPLORER system subsequently tested several of those opportunities in human imaging.