Developed guidelines on quantitative harmonization for oncologic FDG PET/CT.

Nuclear medicine and molecular imaging ORCID OpenAlex

VU University Medical Centre (2015)

Key works · 1

Key2015European Journal of Nuclear Medicine and Molecular Imaging3,675 citations

FDG PET/CT: EANM procedure guidelines for tumour imaging: version 2.0

Ronald Boellaard, Roberto Delgado-Bolton … Wolfgang Eschner

The EANM updated its guidance for adult oncologic FDG PET/CT, covering preparation, acquisition, reconstruction, quality control, quantitation, and reporting. It specified procedures for repeatability on one scanner and for comparability across instruments and sites, including harmonization of quantitative performance across imaging centers.

SignificanceThe guideline specifies procedures and equipment checks for quantitative comparability. The EANM's EARL accreditation report documented implementation across about 200 scanners at 150 sites. Phantom measurements identified calibration errors, and corrective feedback brought participating scanners into compliance with the calibration specifications.