point spread functionPSF
点扩散函数
The image a system makes of a single ideal point: the blur spot a point is smeared into. It sums up how blurry the system is; the narrower the spot, the better the spatial resolution.
If a system is linear and behaves the same everywhere, the whole image is the true object convolved with the PSF, plus noise. Its width is usually quoted as the full width at half maximum (FWHM); the magnitude of its Fourier transform is the modulation transfer function, which says how much contrast survives at each spatial frequency.
The idea comes from optics. In 1835 the astronomer George Airy calculated the diffraction pattern a circular lens makes of a point of light: a bright core surrounded by faint rings, now called the Airy disc. It is the PSF of an ideal lens.
In medical imaging the PSF directly affects how small lesions are depicted. A lesion smaller than two or three times the FWHM has its signal spread thin and looks fainter than it is, the partial-volume effect behind PET's underestimation of SUV in small lymph nodes. Modern PET reconstruction builds the measured PSF into the algorithm and recovers part of the lost resolution.
- Airy GB. On the diffraction of an object-glass with circular aperture. Trans Cambridge Philos Soc. 1835;5:283–291.