ALARA
合理可行尽量低原则
"As low as reasonably achievable": keep radiation dose as low as practical while still answering the clinical question. It is why CT settings are tailored to body size and to children, and why the goal is a diagnostic image, not the prettiest one.
The wording evolved. The ICRP asked in 1959 for doses "as low as practicable", in 1966 "as low as readily achievable", and in Publication 22 (1973) "as low as reasonably achievable, economic and social considerations being taken into account". Publication 26 in 1977 made it the principle of optimization, alongside justification and dose limits.
In a radiology department it becomes concrete: tube current and voltage matched to the patient's size, automatic exposure control, a scan range no longer than needed, dedicated pediatric protocols. The Image Gently campaign, launched in 2008, focused attention on children.
The other side of ALARA is image quality. Halve the dose and quantum noise rises by about ; a lesion that can no longer be seen may mean a repeat scan and more dose, not less. The principle is a balance between signal-to-noise ratio and risk, and low-dose CT protocols and newer reconstruction methods keep moving that balance point.
- Clarke RH, Valentin J. The history of ICRP and the evolution of its policies. Ann ICRP. 2009;39(1):75–110. www.icrp.org/docs/The%20History%20of%20ICRP%20and%20the%20Evolut
- ICRP. The 2007 Recommendations of the International Commission on Radiological Protection. ICRP Publication 103. Ann ICRP. 2007;37(2–4). www.nrc.gov/docs/ml1233/ML12338A682.pdf
- Image Gently Alliance (Alliance for Radiation Safety in Pediatric Imaging). www.imagegently.org/
Key works · 3
Landmark2007Annals of the ICRP
The 2007 Recommendations of the International Commission on Radiological Protection
The Commission revised the radiation and tissue weights used for equivalent and effective dose and defined planned, emergency, and existing exposure situations. It applied justification and optimization across all three situations and specified individual dose limits for occupational and public exposure in planned situations.
SignificanceThe recommendations supplied the framework used in the IAEA International Basic Safety Standards. Those standards apply the principles of justification and optimization to medical exposures and assign responsibilities for implementing them in diagnostic and interventional imaging.
DOI Jack Valentin radiation dosesievertALARAionizing radiation
Key2017Annals of the ICRP1,338 citations
ICRP Publication 135: Diagnostic Reference Levels in Medical Imaging
The Commission clarified which quantities to use for diagnostic reference levels, how to survey practices, and when to update benchmark values. It extended practical guidance to interventional procedures, pediatric imaging, and newer technologies and discussed how automated dose reporting could support periodic comparisons.
SignificanceThe report turns optimization into a process that can be audited across facilities using comparable examinations. The IAEA uses the diagnostic reference level approach in its implementation guidance to identify protocols whose higher typical doses warrant investigation.
DOI Eliseo VañóA. Rogers radiation doseALARAdiagnostic reference levelCT dose index
Key2023Annals of the ICRP59 citations
Optimisation of Radiological Protection in Digital Radiology Techniques for Medical Imaging
The Commission describes how to optimize protection in digital radiography, fluoroscopy, and CT while preserving image quality adequate for the clinical task. It proposes four levels for facilities to assess their practices, covering equipment selection and monitoring, protocol review, staff training, and collaboration across professions.
SignificancePublication 154 gives digital imaging departments a way to assess their implementation of the Commission's optimization principle in stages. The task group's companion strategy describes how departments can use professional roles, technical capability, and management processes to identify their next steps.
DOI Full text Colin J. MartinJenia Vassileva radiation doseALARAdigital radiographyfluoroscopycomputed tomographydiagnostic reference level