MG·LB · Researchers

Frederik Maes

Validated mutual information for rigid registration of CT, MRI, and PET.

Registration, atlases and open toolkits ORCID OpenAlex

KU Leuven

Key works · 1

Landmark1997IEEE Transactions on Medical Imaging5,026 citations

Multimodality image registration by maximization of mutual information

F. Maes, A. Collignon … P. Suetens

The authors aligned CT, MR, and PET images by maximizing the mutual information of corresponding voxel intensities under rigid transformations. Tests against stereotactic reference alignments showed that subvoxel accuracy could be achieved. Additional experiments explored interpolation, optimization, partial overlap, and image degradation.

SignificanceThe paper provided early reference-based validation of intensity-driven multimodal registration using a stereotactic comparison. Its use of mutual information became a methodological reference, reflected in the later survey of similarity measures, interpolation, optimization, and modality combinations.