Image analysis

mutual informationMI

互信息

A measure of how much information one image's intensities carry about another's, without requiring them to match. That makes it the standard similarity measure for multimodal image registration of CT, MRI, and PET, whose brightness rules are entirely different.

Mutual information comes from Shannon's 1948 information theory: , where is entropy. The intensities at matching positions in two images are paired and counted in a joint histogram. When the images are aligned, bone on CT always meets the same kind of MRI signal, the joint histogram gathers into a few tight clusters, joint entropy is low, and mutual information is high. When they are misaligned, the pairs scatter. Registration searches for the transformation that maximizes .

Two groups proposed the idea independently in 1995: Paul Viola and William Wells at MIT, and André Collignon, Frederik Maes, and colleagues at KU Leuven in Belgium. Both journal papers appeared in 1997. Normalized mutual information, introduced by Colin Studholme and colleagues in 1999, is less sensitive to how much the images overlap.

It made automatic fusion of CT with MRI, or PET with CT, an everyday tool, and it remains a standard metric for multimodal registration in software such as ITK and 3D Slicer.

Sources
  1. Viola P, Wells WM III. Alignment by maximization of mutual information. Int J Comput Vis. 1997;24(2):137–154. doi.org/10.1023/A:1007958904918
  2. Maes F, Collignon A, Vandermeulen D, Marchal G, Suetens P. Multimodality image registration by maximization of mutual information. IEEE Trans Med Imaging. 1997;16(2):187–198. doi.org/10.1109/42.563664

Key works · 7

Landmark1997International Journal of Computer Vision3,011 citations

Alignment by Maximization of Mutual Information

Paul Viola, William M. Wells III

The authors estimated alignment by maximizing mutual information between paired measurements from a model and image. Their implementation estimated probability densities from samples and optimized the transformation stochastically. Experiments included MR registration and object alignment in scenes containing clutter or occlusion, where the imaging relation can vary.

SignificanceThe criterion allowed alignment to exploit statistical dependence across intensities measured under different imaging conditions. Subsequent medical registration surveys identify this line of work among the origins of mutual-information registration and document its use across imaging modalities.

Landmark1997IEEE Transactions on Medical Imaging5,026 citations

Multimodality image registration by maximization of mutual information

F. Maes, A. Collignon … P. Suetens

The authors aligned CT, MR, and PET images by maximizing the mutual information of corresponding voxel intensities under rigid transformations. Tests against stereotactic reference alignments showed that subvoxel accuracy could be achieved. Additional experiments explored interpolation, optimization, partial overlap, and image degradation.

SignificanceThe paper provided early reference-based validation of intensity-driven multimodal registration using a stereotactic comparison. Its use of mutual information became a methodological reference, reflected in the later survey of similarity measures, interpolation, optimization, and modality combinations.

Landmark1999IEEE Transactions on Medical Imaging5,930 citations

Nonrigid registration using free-form deformations: application to breast MR images

D. Rueckert, L.I. Sonoda … D.J. Hawkes

The method modeled global breast motion with an affine transform and local deformation with B-spline control points. It optimized normalized mutual information with a smoothness penalty. In three-dimensional contrast-enhanced MRI from volunteers and patients, it recovered deformation more effectively than the rigid and affine registrations tested.

SignificanceThe work demonstrated a practical combination of a B-spline control grid and a voxel-based matching criterion. B-spline free-form deformation became a reference transformation family in medical registration and is configurable in modular tools such as elastix.

Key1999Pattern Recognition2,434 citations

An overlap invariant entropy measure of 3D medical image alignment

C. Studholme, D.L.G. Hill, D.J. Hawkes

The paper addressed changes in overlap as three-dimensional medical images move relative to one another. It proposed a similarity measure given by the sum of marginal entropies divided by joint entropy. Experiments with a simple image model and clinical data showed improved behavior over varying fields of view.

SignificanceNormalized mutual information became a way to address sensitivity to overlap in entropy-based registration. The breast MRI free-form deformation study incorporated this measure into a spatially varying registration model designed for images whose intensities change after contrast administration.

Key2003IEEE Transactions on Medical Imaging3,305 citations

Mutual-information-based registration of medical images: a survey

J.P.W. Pluim, J.B.A. Maintz, M.A. Viergever

The review explained entropy and mutual information, then classified registration methods by choices such as preprocessing, interpolation, optimization, similarity variant, and transformation model. It organized applications by anatomy and modality and summarized comparative studies of the available techniques on medical images.

SignificanceThe review gave researchers a common vocabulary for comparing mutual-information implementations and their applications. The broader survey of deformable registration by Sotiras and colleagues refers readers to this work for detailed coverage of mutual-information methods.

Key2010IEEE Transactions on Medical Imaging4,869 citations

elastix: A Toolbox for Intensity-Based Medical Image Registration

S. Klein, M. Staring … J. Pluim

The paper described elastix, a toolbox that separates the transformation, similarity metric, optimizer, and other registration components into configurable modules. It provided a command-line interface for scripted processing and illustrated how varying individual components affected registration in three example experiments.

SignificanceThe toolbox made it possible to store an entire registration configuration and reuse it across scans or experimental comparisons. SimpleElastix later exposed elastix through SimpleITK language bindings, extending the same modular registration workflows to additional programming environments.

Key2011NeuroImage5,344 citations

A reproducible evaluation of ANTs similarity metric performance in brain image registration

Brian B. Avants, Nicholas J. Tustison … James C. Gee

The study compared mutual information, cross-correlation, and mean squares within the open ANTs registration framework. It evaluated propagated anatomical labels on shared brain MRI data to assess the choice of metric during affine and deformable alignment and described the settings needed to repeat the comparisons.

SignificanceThe paper coupled a shared software implementation with documented metric comparisons and accessible imaging data. ANTs lists it among the citations for the toolkit, allowing published workflows to point to a reference for how registration settings and similarity measures were evaluated.