Modalities · 1980

digital subtraction angiographyDSA

数字减影血管造影

A mask image is taken before contrast and further images as the contrast flows in; subtracting the mask cancels bone and soft tissue, leaving only the opacified vessels. Vascular detail becomes stark, but any movement between frames leaves misregistration artifacts.

Because X-ray attenuation is exponential, the images are log-transformed before subtraction: . Bone, bowel gas and soft tissue are identical in both and cancel; what remains depends only on the iodine along each ray, so even dilute contrast shows clearly.

Subtraction was done on film from the 1930s, laboriously, in the darkroom. In the late 1970s Charles Mistretta's group at the University of Wisconsin, a team at the University of Arizona and another in Kiel, Germany, independently digitized fluoroscopic video and subtracted it in real time; prototype clinical systems were installed in 1980.

Everything depends on nothing moving between frames. Swallowing, breathing and peristalsis leave misregistration artifacts, corrected by shifting or warping the mask, a simple form of image registration. Rotational DSA now reconstructs vessels in 3D and is a principal tool in neurointervention.

Sources
  1. US Congress, Office of Technology Assessment. Introduction to digital subtraction angiography (Health Technology Case Study). 1985. www.princeton.edu/~ota/disk2/1985/8506/850605.PDF
  2. Kruger RA, Mistretta CA, Houk TL, et al. Computerized fluoroscopy in real time for noninvasive visualization of the cardiovascular system. Radiology. 1979;130:49–57. doi.org/10.1148/130.1.49

Key works · 2

Key1979Investigative Radiology72 citations

Computerized Fluoroscopy Techniques for Intravenous Study of Cardiac Chamber Dynamics

Robert A. Kruger, Charles A. Mistretta … David L. Ergun

The group used a computerized fluoroscopy system to extract signals from a peripheral injection of contrast and study ventricular wall motion in dogs and humans. Two time-dependent subtraction algorithms were illustrated in dogs with experimentally induced infarcts, including one that displayed abnormal motion as differences in gray level.

SignificanceThe study demonstrated that digital time subtraction could make cardiovascular contrast visible on fluoroscopic images. The Wisconsin group subsequently used the approach to show carotid, abdominal, and peripheral vessels, documenting early clinical applications of digital subtraction angiography.

Key1980AJR American Journal of Roentgenology156 citations

Computerized Fluoroscopy: Digital Subtraction for Intravenous Angiocardiography and Arteriography

Andrew B. Crummy, Charles M. Strother … Robert A. Kruger

The system used digital time and energy subtraction to enhance the iodine signal from image-intensified fluoroscopy. The authors presented clinical examples of time subtraction after intravenous contrast injection, showing the heart and carotid, abdominal, and peripheral arteries. The examples included stenoses, occlusions, and aneurysms.

SignificanceThe images extended the group's earlier work on cardiac motion to clinical examples of vascular disease in several arterial territories. The team subsequently reported arterial imaging with the same intravenous subtraction approach, documenting where computerized fluoroscopy could be used for vascular assessment.