MG·LB · Key works

Radiography, fluoroscopy and contrast imaging

In the exhibits1.1 1895: Röntgen's discovery of X-rays1.2 Production of X-rays1.3 Attenuation and tissue contrast1.4 Projection imaging and its limits1.5 Contrast agents and specialized radiography1.6 Early use and radiation protection7.2 Digital radiographs and spiral CT9.4 Dual-source CT, tomosynthesis and PET/MR

Landmark1896Science821 citations

On a New Kind of Rays

W. C. Röntgen

This English translation presents the experiments Röntgen reported in 1895. A fluorescent screen lit up near a discharge tube even when black paper blocked the tube's visible light. Objects interposed between the tube and the screen showed different degrees of transmission, and the bones in a hand cast distinct shadows.

SignificanceThe observations established the physical basis of radiography. Radiation could pass through opaque objects, and differences in transmission through their internal structures could be recorded on a screen or photographic plate. The discovery earned Röntgen the inaugural Nobel Prize in Physics in 1901.

Landmark1913Physical Review164 citations

A Powerful Röntgen Ray Tube with a Pure Electron Discharge

W. D. Coolidge

Coolidge described an X-ray tube in which a heated tungsten filament supplies electrons in a high vacuum. A separate high voltage accelerates the electrons toward the target. The design let the operator regulate tube current through filament heating and control the energy of the electrons through the accelerating voltage.

SignificanceThe hot cathode and high vacuum made X-ray generation more reproducible and gave operators separate control over tube current and voltage. The IAEA diagnostic radiology physics handbook describes thermionic emission as the operating principle of modern diagnostic X-ray tubes.

Landmark1915United States Patent Office

Method of and apparatus for projecting Röntgen images

Gustav Bucky

Bucky proposed putting an absorbing grid between the object and the photographic plate to intercept secondary radiation. The openings were aligned with rays from the source, allowing the primary beam to reach the plate while preferentially blocking rays traveling at other angles.

SignificanceThe grid established a way to improve radiographic contrast by rejecting scattered radiation before it reached the receptor. Potter's subsequent use of moving parallel strips reduced the visibility of the grid on the radiograph and led to the Potter–Bucky grid used in clinical systems.

Followed byPotter 1921

Key1921Archives of Radiology and Electrotherapy5 citations

The Bucky Diaphragm Principle Applied to Roentgenography

Hollis E. Potter

Potter replaced the crosswise structure of Bucky's grid with parallel metal strips and moved the grid across the plate during exposure. Uniform motion spread the strip shadows across the plate while the strips still intercepted scattered radiation. Spine radiographs illustrate the resulting suppression of grid lines.

SignificanceThe moving strips addressed a practical obstacle to routine scatter rejection in projection radiography. The Science Museum Group documents the commercial introduction of the Potter–Bucky device in 1921 and holds an example of the equipment used in radiology.

Builds onBucky 1915

Landmark1923Physical Review1,936 citations

A Quantum Theory of the Scattering of X-rays by Light Elements

Arthur H. Compton

Compton modeled the scattering of X-rays by light elements as an exchange of energy and momentum between one radiation quantum and one electron. The model predicted a wavelength shift that depended on the scattering angle, in agreement with observations including the shifted line from graphite.

SignificanceCompton received the 1927 Nobel Prize in Physics for this effect. Compton scattering accounts for much of the photon attenuation in soft tissue at diagnostic energies and also sends off-axis photons toward the detector, where they reduce radiographic contrast.

Landmark1948Radiology95 citations

Fluoroscopic Image Brightening by Electronic Means

John W. Coltman

Coltman addressed the low brightness of conventional fluoroscopic screens by converting the X-ray pattern into an electron image. An electric field accelerated the electrons before they struck an output phosphor, adding energy to the image and yielding a brighter display at the same input exposure.

SignificanceElectronic amplification formed the basis of fluoroscopic image intensifiers. The AAPM museum documents how later systems combined it with image minification to increase gain further, eventually allowing the operator to view fluoroscopic images under ordinary room lighting.

Landmark1951American Journal of Roentgenology and Radium Therapy189 citations

Diagnosis of Tumors of the Breast by Simple Roentgenography: Calcifications in Carcinomas

Raúl Leborgne

Leborgne described the appearance and distribution of fine calcifications in breast carcinomas, comparing radiographs with pathological findings. His technique paired an overview radiograph with a small-field exposure over a suspected lesion. The reduced field limited scatter and let the reader inspect finer detail.

SignificanceThe paper helped establish a radiographic feature that could be interpreted against breast pathology. The morphology and distribution of microcalcifications became part of routine mammographic interpretation and can draw attention to lesions that have yet to produce a palpable mass.

Landmark1953Acta Radiologica3,839 citations

Catheter Replacement of the Needle in Percutaneous Arteriography: A New Technique

Sven Ivar Seldinger

Seldinger described a method of placing a flexible catheter through the arterial entry site made by a puncture needle. A guidewire maintained access during the exchange. The report illustrated how the catheter could then be positioned to deliver contrast at a selected point in the circulation.

SignificanceMaintaining access with a guidewire made selective arteriography possible through a percutaneous puncture. The named Seldinger technique became a standard method of vascular access for diagnostic angiography and for a wide range of interventions performed through a catheter.

Key1966Circulation582 citations

Selective Cine Coronary Arteriography: Correlation with Clinical Findings in 1,000 Patients

William L. Proudfit, Earl K. Shirey, F. Mason Sones, Jr.

The authors reviewed 1,000 patients with adequate selective cine coronary arteriograms. A physician assessed the clinical records without seeing the angiograms. Significant obstruction appeared in 95% of those diagnosed with angina without rest pain. Agreement with the angiographic findings was lower for less characteristic syndromes.

SignificanceThis series showed what selective cine arteriography could add to clinical assessment by documenting coronary obstruction where symptoms gave an uncertain indication. The same group used the clinical and angiographic comparison in subsequent work to study how presenting features related to the extent of obstructive lesions.

Key1972Radiology70 citations

X-Ray Exposure Reduction Using Rare-Earth Oxysulfide Intensifying Screens

Robert A. Buchanan, Solon I. Finkelstein, Kenneth A. Wickersheim

This study compared terbium-activated gadolinium and lanthanum oxysulfide screens with calcium tungstate. The rare-earth phosphors absorbed more incident radiation over part of the diagnostic energy range and converted it to light more efficiently. A gadolinium oxysulfide screen reduced exposure at comparable resolution even with blue-sensitive film.

SignificanceMatching rare-earth screens to green-sensitive film let radiographic systems use more of the light produced per exposure. UNSCEAR records faster rare-earth screens among the technical advances that reduced patient doses in conventional radiography during the screen-film era.

Key1979Investigative Radiology72 citations

Computerized Fluoroscopy Techniques for Intravenous Study of Cardiac Chamber Dynamics

Robert A. Kruger, Charles A. Mistretta, Theodore L. Houk, Wayne Kubal, Stephen J. Riederer, David L. Ergun

The group used a computerized fluoroscopy system to extract signals from a peripheral injection of contrast and study ventricular wall motion in dogs and humans. Two time-dependent subtraction algorithms were illustrated in dogs with experimentally induced infarcts, including one that displayed abnormal motion as differences in gray level.

SignificanceThe study demonstrated that digital time subtraction could make cardiovascular contrast visible on fluoroscopic images. The Wisconsin group subsequently used the approach to show carotid, abdominal, and peripheral vessels, documenting early clinical applications of digital subtraction angiography.

Followed byCrummy 1980

Key1980AJR American Journal of Roentgenology156 citations

Computerized Fluoroscopy: Digital Subtraction for Intravenous Angiocardiography and Arteriography

Andrew B. Crummy, Charles M. Strother, Joseph F. Sackett, David L. Ergun, Chorng G. Shaw, Robert A. Kruger

The system used digital time and energy subtraction to enhance the iodine signal from image-intensified fluoroscopy. The authors presented clinical examples of time subtraction after intravenous contrast injection, showing the heart and carotid, abdominal, and peripheral arteries. The examples included stenoses, occlusions, and aneurysms.

SignificanceThe images extended the group's earlier work on cardiac motion to clinical examples of vascular disease in several arterial territories. The team subsequently reported arterial imaging with the same intravenous subtraction approach, documenting where computerized fluoroscopy could be used for vascular assessment.

Builds onKruger 1979

Landmark1983Radiology978 citations

Computed radiography utilizing scanning laser stimulated luminescence.

M. Sonoda, M. Takano, J. Miyahara, H. Kato

The Fuji system recorded a latent X-ray image on a storage phosphor plate. A laser subsequently scanned the plate and released light in proportion to the stored signal, which the reader converted to digital values. The plate could then be erased and reused.

SignificanceStorage plates made it possible to digitize projection radiographs while retaining cassette-based acquisition in existing rooms. Digital processing and archiving could therefore be introduced before flat-panel detectors were widely available. The IAEA handbook describes this technology as computed radiography.

Landmark1985The Lancet1,956 citations

Reduction in mortality from breast cancer after mass screening with mammography: Randomised Trial from the Breast Cancer Screening Working Group of the Swedish National Board of Health and Welfare

L. Tabár, A. Gad, L. H. Holmberg, U. Ljungquist, Kopparberg County Project Group, C. J. G. Fagerberg

The Swedish Two-County Trial randomized communities to receive or forgo invitations to periodic single-view mammography. Among 134,867 women aged 40 to 74 at entry, the group invited for screening had 31% lower breast cancer mortality and a 25% lower rate of stage II or more advanced breast cancers through 1984.

SignificanceThe two-county trial provided randomized evidence based on breast cancer mortality as well as tumor detection. The International Agency for Research on Cancer used this trial in its assessment of whether population mammography screening could reduce deaths from breast cancer.

Key1990Radiology1,612 citations

Adverse reactions to ionic and nonionic contrast media. A report from the Japanese Committee on the Safety of Contrast Media.

H. Katayama, K. Yamaguchi, T. Kozuka, T. Takashima, P. Seez, K. Matsuura

This prospective nationwide comparison in Japan recorded adverse reactions in 337,647 contrast examinations. Reactions were less frequent with low-osmolar nonionic agents than with high-osmolar ionic agents. Severe reactions occurred in 0.04% of examinations using nonionic agents and in 0.22% of those using ionic agents.

SignificanceThe size of the study allowed a comparison of uncommon severe reactions across contrast formulations and supported the move toward low-osmolar agents. The American College of Radiology cites the study in its contrast media manual when discussing acute reactions.

Key1995NIST2,156 citations

Tables of X-Ray Mass Attenuation Coefficients and Mass Energy-Absorption Coefficients 1 keV to 20 MeV for Elements Z = 1 to 92 and 48 Additional Substances of Dosimetric Interest

J. H. Hubbell, S. M. Seltzer

The NIST report compiled photon mass attenuation and mass energy-absorption coefficients over energies from 1 keV to 20 MeV. The tables and graphs cover elements with atomic numbers 1 through 92 and 48 additional substances and mixtures relevant to dosimetry.

SignificanceThe compilation extended Hubbell's earlier reference tables and was subsequently made available as NIST Standard Reference Database 126. It provides consistent material data for comparing tissue attenuation, calculating photon penetration in shielding, and estimating energy deposition in dosimetry.

Key1995Medical Physics355 citations

X-ray imaging using amorphous selenium: Feasibility of a flat panel self-scanned detector for digital radiology

Wei Zhao, J. A. Rowlands

Zhao and Rowlands assessed a flat-panel concept that used amorphous selenium to convert X-rays directly to electrical charge and a thin-film transistor matrix to read out the resulting image. The analysis indicated feasibility for projection radiography, with electronic noise still limiting performance at typical fluoroscopic exposure rates.

SignificanceThe analysis set design requirements for a direct-conversion radiographic panel. The same team subsequently reported images of a hand phantom obtained with an operating prototype. Later clinical reports document use of amorphous-selenium flat panels for routine projection radiography.

Landmark1997Radiology846 citations

Digital Tomosynthesis in Breast Imaging

L. T. Niklason, B. T. Christian, L. E. Niklason, D. B. Kopans, D. E. Castleberry, B. H. Opsahl-Ong

The investigators acquired multiple low-dose projections on a full-field digital mammography system as the X-ray source moved through an arc. They reconstructed sections from a single arc and tested them on phantoms and mastectomy specimens. Readers generally preferred the sections for lesion visualization at a dose comparable to single-view mammography.

SignificanceThe work demonstrated that a digital breast system could acquire the limited-angle data needed to reduce tissue superposition. The same acquisition principle became the basis of clinical breast tomosynthesis and was subsequently evaluated against conventional digital mammography in screening studies.

Key1997Medical Physics120 citations

Empirical investigation of the signal performance of a high-resolution, indirect detection, active matrix flat-panel imager (AMFPI) for fluoroscopic and radiographic operation

L. E. Antonuk, Y. El-Mohri, J. H. Siewerdsen, J. Yorkston, W. Huang … R. A. Street

This study measured the behavior of a large-area active-matrix flat panel in radiographic and fluoroscopic modes. An intensifying screen converted X-rays to light, which a matrix of amorphous-silicon photodiodes and thin-film transistors read out. The authors characterized dark signals, temporal response, and sensitivity with different phosphor screens.

SignificanceThe measurements showed which signal and timing properties must be managed to use a single indirect panel for still radiographs and successive fluoroscopy frames. Scintillators coupled to amorphous-silicon readout arrays subsequently became a common flat-panel detector architecture.

Notable1999Radiology212 citations

Principles of Digital Radiography with Large-Area, Electronically Readable Detectors: A Review of the Basics

Harrell G. Chotas, James T. Dobbins III, Carl E. Ravin

The review compared large-area electronic X-ray detectors. It explained indirect conversion using a scintillator and light-sensitive elements, direct conversion using charge generated in a photoconductor, and how the detector array reads out the resulting signal. It also discussed implications for image quality.

SignificanceThe article documents the transition from systems requiring separate plate readout to electronically readable area detectors. Its distinction between direct and indirect conversion remains a framework used in medical physics teaching to explain how digital radiographic detectors collect signals.

Key2005New England Journal of Medicine2,187 citations

Diagnostic Performance of Digital versus Film Mammography for Breast-Cancer Screening

Etta D. Pisano, Constantine Gatsonis, Edward Hendrick, Martin Yaffe, Janet K. Baum, Suddhasatta Acharyya

In DMIST, 49,528 asymptomatic women underwent both digital and film screening mammography. Readers interpreted the images independently, and complete data were available for 42,760 women. Receiver operating characteristic analysis showed similar overall accuracy, with higher accuracy for digital mammography among women under 50, women with dense breasts, and premenopausal or perimenopausal women.

SignificanceThe paired design established a direct comparison during the transition from film to digital screening. The National Cancer Institute uses DMIST when summarizing the evidence on digital mammography, including the observed differences by age and breast density.

Landmark2007Annals of the ICRP

The 2007 Recommendations of the International Commission on Radiological Protection

International Commission on Radiological Protection

The Commission revised the radiation and tissue weights used for equivalent and effective dose and defined planned, emergency, and existing exposure situations. It applied justification and optimization across all three situations and specified individual dose limits for occupational and public exposure in planned situations.

SignificanceThe recommendations supplied the framework used in the IAEA International Basic Safety Standards. Those standards apply the principles of justification and optimization to medical exposures and assign responsibilities for implementing them in diagnostic and interventional imaging.

Followed byProtection 2023

Landmark2011New England Journal of Medicine12,304 citations

Reduced Lung-Cancer Mortality with Low-Dose Computed Tomographic Screening

National Lung Screening Trial Research Team

At 33 centers, the National Lung Screening Trial randomly assigned high-risk participants to three annual rounds of low-dose CT or single-view chest radiography. Lung cancer mortality was 20% lower in the CT group. That group also had more positive screening results, most of which were false positives.

SignificanceThe trial showed that the evaluation of low-dose CT screening could measure lung cancer deaths along with detection and false-positive results. The US Preventive Services Task Force included NLST in its evidence for lung cancer screening recommendations.

Key2014JAMA927 citations

Breast Cancer Screening Using Tomosynthesis in Combination With Digital Mammography

Sarah M. Friedewald, Elizabeth A. Rafferty, Stephen L. Rose, Melissa A. Durand, Donna M. Plecha, Julianne S. Greenberg

The investigators retrospectively compared 454,850 examinations across 13 centers before and after adding tomosynthesis to digital mammography. In adjusted estimates per 1,000 screens, recall declined from 107 to 91 and cancer detection increased from 4.2 to 5.4 after the addition.

SignificanceThe study estimated changes in recall and detection across a range of screening practices after clinical adoption. It was included in the Agency for Healthcare Research and Quality evidence review and supplied multicenter data on outcomes later evaluated in prospective trials.

Key2017Annals of the ICRP1,338 citations

ICRP Publication 135: Diagnostic Reference Levels in Medical Imaging

E. Vañó, D. L. Miller, C. J. Martin, M. M. Rehani, K. Kang, M. Rosenstein

The Commission clarified which quantities to use for diagnostic reference levels, how to survey practices, and when to update benchmark values. It extended practical guidance to interventional procedures, pediatric imaging, and newer technologies and discussed how automated dose reporting could support periodic comparisons.

SignificanceThe report turns optimization into a process that can be audited across facilities using comparable examinations. The IAEA uses the diagnostic reference level approach in its implementation guidance to identify protocols whose higher typical doses warrant investigation.

Key2022The Lancet Oncology139 citations

Digital breast tomosynthesis plus synthesised mammography versus digital screening mammography for the detection of invasive breast cancer (TOSYMA): a multicentre, open-label, randomised, controlled, superiority trial

Walter Heindel, Stefanie Weigel, Joachim Gerß, Hans-Werner Hense, Alexander Sommer … TOSYMA Screening Trial Study Group

Across 17 German screening units, TOSYMA randomly assigned women to either digital breast tomosynthesis with synthesized two-dimensional mammography or standard digital mammography. Invasive cancers were detected at rates of 7.1 and 4.8 per 1,000 screened, respectively, in the evaluable groups.

SignificanceTOSYMA provided randomized evidence for higher invasive cancer detection with a configuration that obtains the two-dimensional image from the same acquisition. It therefore evaluated a screening protocol that can be implemented without a separate conventional two-dimensional exposure.

Followed byWeigel 2026

Key2023Annals of the ICRP59 citations

Optimisation of Radiological Protection in Digital Radiology Techniques for Medical Imaging

International Commission on Radiological Protection

The Commission describes how to optimize protection in digital radiography, fluoroscopy, and CT while preserving image quality adequate for the clinical task. It proposes four levels for facilities to assess their practices, covering equipment selection and monitoring, protocol review, staff training, and collaboration across professions.

SignificancePublication 154 gives digital imaging departments a way to assess their implementation of the Commission's optimization principle in stages. The task group's companion strategy describes how departments can use professional roles, technical capability, and management processes to identify their next steps.

Builds onProtection 2007

Notable2025JAMA Network Open72 citations

Efficiency and Quality of Generative AI–Assisted Radiograph Reporting

Jonathan Huang, Matthew T. Wittbrodt, Caitlin N. Teague, Eric Karl, Galal Galal, Michael Thompson

Northwestern Medicine phased in generated radiograph report drafts in its live workflow. In the prospective cohort, use was associated with 15.5% shorter documentation time compared with a pre-use baseline matched by exam type. Peer review of final radiologist-edited reports from 800 studies found no significant clinical accuracy difference.

SignificanceThe prospective cohort documents generated drafts in routine imaging within one health system. Documentation time and final report quality were assessed in separate groups of exams from the same workflow. The timing comparison estimates an association.

Key2026The Lancet Oncology1 citations

Invasive interval breast cancer rates after digital breast tomosynthesis plus synthesised mammography versus digital mammography screening (TOSYMA): 24-month follow-up of the multicentre, open-label, randomised, controlled, superiority trial

Stefanie Weigel, Hans-Werner Hense, Veronika Weyer-Elberich, Jörg Czwoydzinski, Thomas Decker, Hiltraud Kajüter

This TOSYMA follow-up linked women with a negative screening result to cancer registries. At 24 months, the tomosynthesis plus synthesized mammography arm had 14.2 invasive interval cancers per 10,000 women, compared with 18.6 with digital mammography. The adjusted odds ratio was 0.76, with a 95% confidence interval of 0.55 to 1.06.

SignificanceThe follow-up reports the second prespecified primary endpoint of the same randomized trial that found higher invasive cancer detection at screening in 2022. The confidence interval leaves uncertainty about whether higher detection at screening translates into fewer invasive interval cancers over the next two years.

Builds onHeindel 2022